Analytical Instrumentation & Environmental Monitoring


Stack gas analyser for hot/wet analysis

November 2000 Analytical Instrumentation & Environmental Monitoring

Ametek's 910 mass emission analyser uses the 920 stack gas analyser for the analysis of SO2. It maintains the sample above the water dewpoint and analyses on a hot/wet basis. An important fact is that the velocity/flow is measured on an in-situ hot/wet basis, and as the mass emission is a product of these two values, the measurement base must be the same.

This method of extractive analysis is commonly referred to as hot/wet analysis. That is, the sample is extracted from the process and kept above the sample dew point throughout the sample transport and analysis. There is no dilution or drying of the sample; keeping the sample above the dewpoint not only refers to water but also to acid dewpoints - the sample system is capable of maintaining a sample temperature of up to 180°C or removing the acid at the SCU. This means sample integrity is maintained and any plugging or fouling is minimised.

The 900 Series analysers use hollow cathode lamps, which emit characteristic lines of UV radiation. The positions of these lines are determined by the cathode metal, their wavelength centres being known to a high degree of accuracy. Since these bands are extremely narrow, they facilitate excellent linearity, accuracy and precision in absorbance measurements, independent of variations in bandpass filtering.

Separate reference and measurement paths are part of the photometer design. Light from the source lamp is split into two separate beams through the use of a beam splitter. One beam passes through clean air, the other through the gas sample. Both beams impinge on separate PMTs (reference and measure).

The sample cell is made of stainless steel, using a folded path configuration, providing for the photometer to be mounted outside of the oven in the electronics cabinet. At one end of the sample cell, a reflector block returns the photometer illumination back to the detector section through a thermal isolation tube. Ametek is represented locally by OEN Enterprises.





Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Getting the chemistry right
Analytical Instrumentation & Environmental Monitoring
Grace’s European production site has relied on analogue compression load cells from Minebea Intec for decades, with the long-proven measurement concept now serving as the reference for new silo projects as the company expands its capacity.

Read more...
Why measurement matters in smelting
Endress+Hauser South Africa Analytical Instrumentation & Environmental Monitoring
South African smelters returning to operation after energy constraints face new demands for tighter control, better visibility and environmental compliance. Endress+Hauser explains how modern measurement technology supports safer, more efficient furnace operation and local beneficiation.

Read more...
Sun tracking with pinpoint precision
Senseca Analytical Instrumentation & Environmental Monitoring
Senseca has introduced the SEMS-2000 solar energy measurement system, a 2-axis solar tracker with pinpoint sun-pointing accuracy that automatically monitors direct, global and diffuse irradiance for unattended outdoor solar installations.

Read more...
AMETEK expands 993X gas analyser series with explosion-proof enclosures
Analytical Instrumentation & Environmental Monitoring
AMETEK Process Instruments has added ATEX and IECEx Zone 1-certified explosion-proof enclosures to its 993X series of gas analysers, enabling reliable installation in Zone 1 hazardous locations where purge gas is unavailable.

Read more...
Why medical device leak testing matters for South Africa's MedTech future
ATEQ South Africa Analytical Instrumentation & Environmental Monitoring
As South Africa works to grow local medical technology manufacturing, non-destructive leak and flow testing is becoming a strategic quality assurance tool, helping manufacturers verify device integrity, meet regulatory requirements and improve production efficiency.

Read more...
Moving forward with reverse osmosis
Analytical Instrumentation & Environmental Monitoring
Rising water costs and pressure on infrastructure are changing the way industries view reverse osmosis. G-Chem Aquacare is developing tailored, containerised solutions that make the technology more accessible for South African industry.

Read more...
Meteorological data logger with up to 32 analogue sensors
Senseca Analytical Instrumentation & Environmental Monitoring
Senseca has introduced its advanced data acquisition system for environmental monitoring, which has a universal data logger for up to 32 analogue, pulse and smart digital sensors.

Read more...
Importance of gas analysis in the petrochemical industry
Elemental Analytics Analytical Instrumentation & Environmental Monitoring
Gas analysis plays a critical role in petrochemical operations, supporting process optimisation, product quality and plant safety. As an experienced and expert gas analysis supplier, Elemental Analytics can provide the correct analyser and sample system package to meet individual plant requirements.

Read more...
Fully integrated standalone thunderstorm warning system
Senseca Sensors & Transducers Analytical Instrumentation & Environmental Monitoring
Senseca South Africa has introduced its BTD-1 fully integrated, standalone thunderstorm warning system, which detects lightning by monitoring the background quasi-electrostatic field. The system can generate alerts before the first lightning strike, with a detection raterange of up to 83 km.

Read more...
Environmental monitoring with testo Saveris 1
Analytical Instrumentation & Environmental Monitoring
The testo Saveris 1 environmental monitoring system has been in continuous use since 2010 at Jungbunzlauer, a biotechnology company in Germany, providing seamless temperature and humidity data for audit-ready, GxP-compliant processes across laboratories, warehouses and production areas.

Read more...









While every effort has been made to ensure the accuracy of the information contained herein, the publisher and its agents cannot be held responsible for any errors contained, or any loss incurred as a result. Articles published do not necessarily reflect the views of the publishers. The editor reserves the right to alter or cut copy. Articles submitted are deemed to have been cleared for publication. Advertisements and company contact details are published as provided by the advertiser. Technews Publishing (Pty) Ltd cannot be held responsible for the accuracy or veracity of supplied material.




© Technews Publishing (Pty) Ltd | All Rights Reserved