How are infrared cameras calibrated?

October 2021 Maintenance, Test & Measurement, Calibration

Knowing the exact value of the radiation temperature is essential for the calibration process. It can be measured either by using a contact thermometer (in combination with the determination of the emissivity) or a transfer standard infrared thermometer. This value can then be used to determine the device constant for the initial calibration of the infrared sensors. The calibration temperature should be near the temperatures which occur at the respective applications.

Instrotech, the local supplier of Optris products, makes use of a transfer standard radiation thermometer CTlaser-PTB to measure the radiation temperature of a reference source. This instrument needs to be traceable to the international temperature scale from 1990 (ITS-90), so it is calibrated regularly by the PTB (German National Metrology Institute).

ITS-90 is a good approximation of thermodynamic temperature. It is based on 17 well-reproducible fixed values such as melting points of highly pure metals. Within the framework of ITS-90 the CTlaser-PTB is compared to national temperature standards from the PTB. This comparison within a closed chain of comparative measurements with a known uncertainty in measurement takes place regularly.

Based on the CTlaser-PTB, Optris produces the CTlaser- DCI as a high-precision reference IR thermometer for its customers. The DCI units are produced with pre-selected components supporting high stability of measurement. In combination with a dedicated calibration at several calibration points, the CTlaser-DCI achieves higher accuracy than units from series production.

Optically, an IR thermometer is described by the distance-to-spot ratio (D:S). Depending on the quality of the optics, a certain amount of radiation is also received from sources outside the specified measurement spot. The maximum value here equals the radiation emitted by a hemispheric radiant source. The respective signal change in correlation with a resize of the radiation source is described by the Size-of-source effect (SSE). As a result of this correlation, manufacturers of IR thermometers use accurately defined geometries for the calibration of their units, meaning that depending on the aperture of the radiation source (A) a distance (a) between the IR thermometer and the reference source is defined. Thus, the value specified in datasheets and technical documentation as measurement field is in general a certain defined percentage of this radiation maximum – values of 90% or 95% are common.

Optris has up-to-date in-house laboratories which fulfil the mandatory requirements for calibration stations. When issuing calibration certificates, it is not only the laboratory temperature and humidity that is documented but also the measurement distance and source diameter (calibration geometry).

Ambient temperature referencing source BR20AR

To improve the specified camera accuracy of the PI 450i T010 camera, a reference source with a high emissivity and a stable and known temperature must be positioned close to the subject to be scanned. The BR 20AR Ambient referencing source is equipped with a temperature probe with 0,1°C accuracy.

By integrating this highly accurate reference signal with the PIX Connect software, camera uncertainties resulting from device adjustment, ambient temperature drift and short-term stability, can be reduced to a system accuracy of 0,5°C. Other BR20AR features include:

• Reference radiator with high emissivity ideal for IR camera-based fever screening applications.

• Integrated 16-bit digital temperature sensor with 0,1 °C accuracy.

• Mounting bracket – adjustable for either ceiling or wall mounting.

• Plug-and-play installation with 20 m cable and suitable PIF connector.


Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Collect data three times faster
SKF South Africa Maintenance, Test & Measurement, Calibration
SKF has extended its renowned Microlog Analyzer family of data collection devices with the addition of the Microlog Analyzer dBX. Currently SKF’s most powerful diagnostic tool, this cutting-edge device redefines diagnostic capabilities, enabling users to take measurements three times faster than its predecessor.

Read more...
The criticality of maintenance in the water and wastewater segment
Schneider Electric South Africa Maintenance, Test & Measurement, Calibration
In a time of water shedding, climate change and ageing infrastructure, the importance of maintenance and support in the water and wastewater segment cannot be overstated.

Read more...
How dry ice blasting is revolutionising the mining industry
Maintenance, Test & Measurement, Calibration
Dry ice blasting has emerged as a game-changing technology for the mining industry, offering a range of advantages that are reshaping traditional cleaning methods.

Read more...
Pressure sensors for hydrogen applications
Instrotech Pressure Measurement & Control
Keller manufactures isolated piezoresistive pressure sensors. The great advantage of the piezoresistive measuring principle is its high stability.

Read more...
Quality test tools save costs at pulp and paper mill
Comtest Editor's Choice Maintenance, Test & Measurement, Calibration
A case study on how preventive maintenance and a few good test tools avoided unnecessary motor replacements at a pulp and paper company.

Read more...
Wind turbines with smart sensors
Instrotech Motion Control & Drives
The wind’s usable kinetic energy increases with the wind speed. Wind turbines, therefore, use kinetic energy to generate torque and then rotational movement. The rotor, consisting of three rotor blades, ...

Read more...
Infrared camera for condition monitoring and early fire detection
Instrotech Temperature Measurement
The Optris PI 640i thermal imager is the smallest measuring VGA infrared camera, with a body-sized volume of 45 x 56 x 90 mm, and weighing only 269 to 340 g, depending on the lens. It counts among the most compact thermal imaging cameras on the market.

Read more...
Pressure monitoring for oil and water tempering machines
Instrotech Pressure Measurement & Control
Temperature plays a significant role in many manufacturing processes. In such cases, a tempering machine is used to maintain the manufacturing process at the desired temperature. Keller supplied PR-21Y piezoresistive pressure transmitters to a client specialising in developing and producing oil and water tempering machines.

Read more...
The logical solution to oil and gas industry corrosion
Maintenance, Test & Measurement, Calibration
The consequences of corrosion in the oil and gas industry can be astronomical. The simple culprit is metal reacting with oxygen and moisture, degrading it back to its natural state.

Read more...
Lube tip: foaming is affected by oil level
Wearcheck Maintenance, Test & Measurement, Calibration
In a circulating system, it is crucial to check the oil level before introducing anti-foam agents to address a foaming problem.

Read more...