Flow Measurement & Control


Improve DP-based flow measurement with a multivariable pressure transmitter

July 2007 Flow Measurement & Control

This article describes a pressure transmitter which offers built-in dynamic compensation for many of the errors and inaccuracies that can occur in differential pressure-based flow measurements.

One of the most common ways of calculating volumetric or mass flow is to measure the differential pressure (DP) across a primary element such as an orifice plate or Pitot tube. However, using only one instrument, the pressure and density (and hence flowing density) have to be assumed as constant in order to convert the DP value to a flow rate.

The DPharp EJX910A multivariable pressure transmitter from Yokogawa, in combination with its primary element, forms a flowmeter that computes standard volumetric or mass flow from measured differential pressure (DP) and flowing density using actual measured pressure and temperature.

Error sources affecting DP flow measurement

Figure 1 shows the measurement set-up. The basic mass flow equation performed is: Qm≈k.√(DP.P/T).

Figure 1. Multivariable flow measurement configuration
Figure 1. Multivariable flow measurement configuration

In DP flow measurements there are several potential error sources. These include:

* Primary element effects.

* DP measurement accuracy.

* Pressure and temperature fluctuations (variation in flowing density).

Primary element effects

The orifice plate is the most widely used primary element, and its accuracy is extensively debated. It is known, however, that its accuracy degrades over time as the plate loses its sharpness. A typical accuracy figure might be 1% of rate.

DP measurement accuracy

The accuracy of the DP transmitter is expressed as percentage of upper range value (URV), and is magnified by the square-law relationship between flow and differential pressure. Traditionally, DP-based flowmeters have featured a 3:1 turndown. Turndown is a figure that expresses the range over which the accuracy is perceived to be acceptable: 30-100% of flow for a 3:1 turndown.

If the accuracy of the DP transmitter is assumed to be ±0,2% URV, at 100% flow the accuracy is ±0,2% plus the error of the plate. At 70% flow (49% DP due to the square-law relationship), the accuracy would be ±0,4%; at 50% flow (25% DP) it would be ±0,8%; and at 25% flow (6,25% DP) it would be ±3,2%. The plate adds ±1% over the full range and, when ±2,5% total accuracy is perceived to be acceptable, that accuracy is reached at around 30% flow - hence the 3:1 turndown. Turndown is often increased to 9:1 by using two DP transmitters over the same plate with different ranges, a switching mechanism and a flow computing function. Today, of course, DP transmitters are much more accurate than ±0,2%. However, the accuracy is still expressed as a percentage of URV, and the square-law relationship between flow and DP is still applicable. Nevertheless, one could argue that a turndown of better than 5:1 is not normally achieved.

Pressure and temperature fluctuations

The error due to pressure (±0,5 bar) and temperature fluctuations (±10°C) in gas or steam flows can easily be of the order of magnitude of ±2%. The flow factor k in the basic formula above is a constant compensating for the differences between theory and real life. It is calculated as part of the orifice calculation, and is valid only for one particular set of operating conditions - making it another potential source of error. It is dependent on the discharge coefficient, the gas expansion factor and velocity of approach factor. The result of an orifice calculation is the diameter of the hole in relation to the pipe diameter: the so-called β-ratio. When a fluid is flowing through a pipe with an orifice plate, the flowing area reduction is abrupt, causing the smallest flowing area (Vena contracta) to be downstream (behind) of the plate. The discharge coefficient (based upon experimental research) compensates for the difference between theory and real life. However, the discharge coefficient is flow profile (Reynolds) dependent and varies with the flow velocity, pipe internal diameter, flowing density and flowing viscosity. In their turn, the latter three parameters are affected by flowing temperature. When a gas or steam flow passes the orifice plate (Figure 2), it is compressed upstream of the plate due to the obstruction caused by the plate. Downstream of the plate it expands again. The gas expansion factor corrects for density differences between the pressure taps, and it depends on the β-ratio, the isentropic coefficient (correcting for theoretical versus real-life expansion), differential pressure and static pressure. Again, temperature is having an effect as well. Finally, the velocity of approach factor is dependent on the β-ratio (d/D), which in turn is dependent on temperature. Pipe and orifice plate material expands or contracts as temperature changes, and the velocity of approach factor corrects for changes in β-ratio due to temperature fluctuation.

Figure 2. Effect of orifice plate on flow
Figure 2. Effect of orifice plate on flow

Dynamic compensation

The error in flow factor k increases with decreasing flow rate, significantly contributing to the overall flowmeter accuracy. A primary element of a DP-based flowmeter is sized for one particular set of operating conditions.

When in real life the conditions change - lower pressure, higher temperature, lower flow rate etc, - the user would like to re-calculate the flow-rate/DP relationship for every new set of conditions. This is exactly what the DPharp EJX910A multivariable transmitter (Figure 3) is able to do. It corrects for pressure and temperature variations and can compensate continuously for the effects of changing operating conditions on the flow factor. Using its flow configuration wizard, the multivariable transmitter can be set up to act as a flow computing device.

Figure 3. The Yokogawa DPharp EJX910A multivariable pressure transmitter
Figure 3. The Yokogawa DPharp EJX910A multivariable pressure transmitter

There are two options: basic mode and auto compensation mode. In basic mode, the transmitter only compensates for pressure and temperature fluctuations with a constant flow factor, in much the same way as a standard DP transmitter with a separate flow computer, resulting in a turndown of perhaps 5:1. However, in the auto compensation mode, the transmitter dynamically compensates for fluctuating conditions and their effects on the flow factor, reducing the error basically to the uncertainty intrinsic to the primary element. Consequently, the accuracy specification of ±1% of flow rate (assuming the primary element to be ideal) over 10:1 turndown in flow terms (equalling 100:1 turndown in DP terms) is only valid in the auto compensation mode.

Conclusion

The multivariable transmitter breathes new life into existing square-law flow measurements and, with its 10:1 turndown, not only eliminates the need for a separate flow computer, but can replace two DP transmitters over the same plate - achieving good performance over a wide flow range.



Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Ultra-compact mass flow meters with advanced transmitters
Emerson Automation Solutions Flow Measurement & Control
Emerson has introduced the Micro Motion G-Series line of Coriolis mass flow and density meters, the most compact dual-tube Coriolis mass flow meters on the market. They offer the same level of quality and reliability as standard designs, but in a much smaller and lighter form factor.

Read more...
Flow measurement transmitters with IO-link interface
Senseca Flow Measurement & Control
Senseca has introduced the enhanced OMNIPLUS range of transmitters that feature an intelligent IO-Link interface.

Read more...
Optimisation of alkaline electrolysers
Samson Controls Flow Measurement & Control
his application study relates to the use of a smart meter valve in alkaline electrolysers, and is based on proven experience with one of the biggest hydrogen electrolyser producers in the world. The project proves that the smart meter valve’s added functionality creates an ideal setup in alkaline electrolysers for reducing cost and improving functionality and system performance.

Read more...
Preventing spills and improving operational efficiency
Flow Measurement & Control
In an industry where every spill carries a hefty price tag, Point Energy Partners found that the manual and low-tech systems at their saltwater disposal sites were a significant problem.

Read more...
Accurate flowmeter for oil & gas
Endress+Hauser South Africa Flow Measurement & Control
Promass Q, the high-tech Coriolis flowmeter from Endress+Hauser, is now also available for larger pipe sizes with maximum flow rates between 850 and 2400 tph.

Read more...
Optimising CIP processes with precision instrumentation
Endress+Hauser South Africa Flow Measurement & Control
Many food and beverage manufacturers encounter challenges with CIP processes that take too long. Have you considered exploring potential solutions to address these issues?

Read more...
How predictive maintenance will help tackle water scarcity
ABB South Africa Flow Measurement & Control
The water industry is facing increasing pressure to optimise treatment processes and address the impending global water crisis. ABB’s global product line manager, Krishna Prashanth suggests how the integration of smart instrumentation and predictive maintenance can ensure the highest quality treatment of water.

Read more...
Flowmeters with faster and more reliable data transmission
ABB South Africa Flow Measurement & Control
ABB has launched the CoriolisMaster and ProcessMaster flowmeters, which can be powered over the same Ethernet cable used for data transmission, eliminating the need for separate cabling.

Read more...
Ice cold optimisation
Turck Banner Southern Africa Flow Measurement & Control
A well-known industrial company in China specialises in the research, development and production of high-tech materials and special industrial furnaces for various fields and applications. To ensure the safe and efficient operation of the furnaces, the flow and temperature of the cooling water and the energy consumption must be constantly monitored and controlled. Turck’s FS+ flow sensors are ideal for this application as they can monitor the temperature of the cooling medium in addition to the flow.

Read more...
Smart sensor for condition monitoring
Yokogawa South Africa Sensors & Transducers
Yokogawa has developed a compact, intelligent, low-power battery-operated wireless sensor, the LoRaWAN, that can be deployed to perform vibrational condition monitoring in the field.

Read more...