Why and how pressure gauges should be calibrated

June 2017 Maintenance, Test & Measurement, Calibration

What is calibration?

Basic pressure gauge calibration is the comparison of measurement values of a unit under test (your gauge) with those of a more accurate calibrated reference instrument. This instrument is normally traceable to National Standards (NMISA).

The term calibration means just the act of the comparison and the recording of noted values on a certificate. It does not normally include corrective adjustment. The outcome of the comparison can result in the following three reports:

• No significant error being noted.

• A significant error being noted but no adjustment made.

• An adjustment made to correct the error to an acceptable level. (When adjustments are made, the before and after values should be indicated.)

The purpose of calibration is to maintain the quality and accuracy of measurement and to ensure the proper working of a particular instrument.

The importance of calibration

Even the highest quality instruments are subject to drift over time, resulting in inaccurate measurements and substandard performance. It is very important that all instruments are calibrated by trained, competent and approved personnel. Calibration intervals and error limits should be defined and records of the calibration results should be kept, maintaining instrument integrity.

Neglecting calibration can result in unplanned production downtime, inferior products, or process quality issues. In addition, allowing instruments that are critical to the process to drift out of specification could potentially result in a risk to employee safety. The cost of calibration is normally insignificant compared to the potential production or injury costs.

When to calibrate

Calibration intervals as specified by the manufacturer are normally followed until the user can assign calibration intervals based on the history of previous calibration results. In addition, calibration is often required with a new instrument, or when an instrument may have been subjected to an unexpected shock or vibration that may have put it out of its specified limits.

Types of pressure gauges

Different types of pressure measurement exist. These include gauge pressure, vacuum, absolute, barometric, and differential pressure.

Gauges are available for each of the above types of pressures, as well as for compound pressure which indicates pressure or vacuum on the same dial. For the calibration information below, we will use gauge pressure, the most common type.

The most common calibration guidelines

There are many factors influencing calibration. Listed below are a few of the most common that may influence the calibration of gauges in the class 0,25%, 1% and 1,6%.

1. Accuracy class

Pressure gauges are manufactured in many different accuracy classes: most commonly as per EN 837 accuracy classes from 0,1% to 0,4% of range. The class indicates the allowable percentage error of the full-scale value of the gauge. For example, if the full-scale value is 1600 kPa and the gauge class is 1,6% then the maximum allowable error of the gauge will be 25,6 kPa, rounded up to 26 kPa. (Error values should always be rounded up, never round down.)

When calibrating, all values reported should be within the class of the gauge. If the class is unknown, the manufacturer should be contacted. For process applications, the following accuracies normally apply: for dial sizes 40 mm and 50 mm, class 2,5%; for 63 mm class 1,6%; and for 100 mm and 160 mm class 1%. For reference test gauges with mirror scales, the norm is 0,25% accuracy. Remember that these values are the percentage value of the full scale, not the percentage value of readings taken anywhere else on the dial.

2. Pressure media and Adiabatic effect

Lower pressure gauges are normally calibrated with air and higher pressure gauges are calibrated with liquid. For pressures below 40 kPa, gas such as nitrogen or alternatively regular air are the preferred media.

Gas is also practical for use for pressures up to 6000 kPa as long as the Adiabatic cooling effect is eliminated. The Adiabatic cooling effect occurs when gas is rapidly pressurised in a closed system. The gas heats up and expands in volume, resulting in a higher pressure reading than after it has been allowed to cool down and stabilise. Readings should only be taken after allowing the gas to cool down and the system to stabilise.

For pressures above 60 kPa water or oil may be used as a pressure media.

3. Pressurising the gauge

The most common pressure gauges consist of a mechanical element, linked to a mechanical movement with screws or rivets. Pressurise and vent the gauge three times before calibration to ensure repeatable values can be obtained. Each time check for drag or pointer stickiness.

4. Reading the pressure

Different dial sizes and different scale ranges result in varying degrees of difficulty when taking pressure readings. All dials have major and minor scale markings and the distance between the resolution marks is determined by the gauge dial size, the range of the gauge and the class of the gauge. (As a rule of thumb, most resolutions can be divided by four if necessary to take a more accurate reading.)

Applied pressure – pointer on segment of the unit under test. Easy to read true value on reference gauge.
Applied pressure – pointer on segment of the unit under test. Easy to read true value on reference gauge.

Standard practice is to apply pressure to the gauge under test until the pointer lines up perfectly on a graduation, and then take the reading on the more accurate reference instrument that should have a better resolution. It is also important to avoid parallax errors when taking readings. Because the pointer is not against the dial plate, looking at an angle at the gauge will result in incorrect readings. Most test gauges have a mirror dial to assist the operator to line up the reflection of the pointer with the actual pointer, making sure that the reading is taken from a perfectly perpendicular angle.

Applied pressure – pointer on segment of the reference gauge. Difficult to read true value on unit under test.
Applied pressure – pointer on segment of the reference gauge. Difficult to read true value on unit under test.

5. Hysteresis

Hysteresis exists in all mechanical pressure gauges and, if required, it must be recorded on the certificate. Hysteresis is the difference in value at the same calibration point but in opposing directions. Take a reading at a point when the pressure rises and again at the same point with the pressure falling. The difference in value will be the hysteresis. Gentle tapping on the gauge to release any friction is considered standard practice and is recommended.

Note: There are more factors to consider when calibrating pressure gauges and the above information addresses only some of the main points.

For more information contact Chris du Plessis, SA Gauge, +27 (0)86 000 7911, [email protected], www.sagauge.com



Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Infrared thermometers for electrical, industrial and HVAC applications
Comtest Maintenance, Test & Measurement, Calibration
: Comtest, distributor of Fluke test and measurement equipment, explains how handheld infrared thermometers are transforming maintenance in electrical, industrial and HVAC environments, with three measurement strategies and a wide range of practical applications.

Read more...
Reverse-engineering legacy industrial gearbox parts
Maintenance, Test & Measurement, Calibration
WEG Africa’s turbine workshop uses 3D scanning technology to reverse-engineer components for legacy gearboxes, turbines and compressors, enabling equipment to run for decades after OEM support and spare parts have dried up.

Read more...
Unlocking ICCP remote monitoring
Omniflex Remote Monitoring Specialists Maintenance, Test & Measurement, Calibration
Cathodic protection keeps buried and submerged pipelines safe from corrosion, but verifying it traditionally means costly, hazardous manual site visits. Omniflex explains how a retrofit module brings remote monitoring to ageing transformer rectifiers without replacing functional infrastructure.

Read more...
Microdieseling – the explosions you never hear
Maintenance, Test & Measurement, Calibration
Microdieseling is a pressure-induced thermal degradation mechanism in which entrained air bubbles are compressed so rapidly that the surrounding oil ignites, producing soot, varnish and sludge. WearCheck explains how it happens, why the damage varies, and how oil analysis detects it.

Read more...
Electrostatic spark discharge – when static strikes back
Wearcheck Maintenance, Test & Measurement, Calibration
WearCheck technical manager, Steven Lumley explains how static electricity accumulates in modern low-conductivity lubricants, why compact system designs and fine filtration make ESD more likely, and how to recognise the degradation it leaves behind.

Read more...
High-speed cameras for crash testing
Horne Technologies Maintenance, Test & Measurement, Calibration
Vision Research has expanded its Phantom C-series with two cameras optimised for crash and safety testing, offering 4 and 8 megapixel sensors rated to 170G, up to 1180 frames per second, and Phantom Academy application training planned for South Africa next year.

Read more...
SKF extends remanufacturing expertise to bearing clearance modification
SKF South Africa Maintenance, Test & Measurement, Calibration
Using its clearance modification expertise, SKF helped a mining equipment OEM salvage spherical roller bearing inventory that would otherwise have been scrapped, confirming a new capability within its remanufacturing services and avoiding redundant procurement.

Read more...
When maintenance-system compliance starts replacing plant reliability
Wearcheck Maintenance, Test & Measurement, Calibration
WearCheck ARC division manager, Annemie Willer argues that the greatest threat to effective maintenance is no longer a lack of technology, but the gradual replacement of technical judgement by CMMS compliance, leaving no one owning the machine.

Read more...
Automatic lubrication systems drive mobile mine machine reliability and uptime
SKF South Africa Maintenance, Test & Measurement, Calibration
SKF’s Lincoln P203 progressive lubrication systems are improving reliability and uptime on mobile machines at a Zambian gold mine, with installation of 45 units completing by October 2026.

Read more...
Nitration: the acid you didn’t see coming
Wearcheck Maintenance, Test & Measurement, Calibration
Nitration is as destructive as oxidation in many engines but far less understood. WearCheck technical manager, Steven Lumley explains how nitrogen oxide gases degrade lubricants, why the process is difficult to detect, and what oil analysis can reveal before damage occurs.

Read more...









While every effort has been made to ensure the accuracy of the information contained herein, the publisher and its agents cannot be held responsible for any errors contained, or any loss incurred as a result. Articles published do not necessarily reflect the views of the publishers. The editor reserves the right to alter or cut copy. Articles submitted are deemed to have been cleared for publication. Advertisements and company contact details are published as provided by the advertiser. Technews Publishing (Pty) Ltd cannot be held responsible for the accuracy or veracity of supplied material.




© Technews Publishing (Pty) Ltd | All Rights Reserved