Calibration vs verification

March 2019 Maintenance, Test & Measurement, Calibration

Maintenance personnel often come across the metrological terms calibration and verification. For some, these two concepts are known and easy to differentiate, for others, they can cause confusion.

Verification made easy

According to the International Vocabulary of Metrology (VIM), the term verification is defined as: “Provision of objective evidence that a given item fulfils specified requirements.”

An interpretation of verification consists of checking calibration results as ‘objective evidence’ to comply with a ‘specified requirement’, such as the Maximum Permissible Error (MPE), defined either by a manufacturer, a legal metrology organisation or an end-user (i.e. process application). This situation is illustrated in Figure 1, where the device’s relative measurement errors obtained by the calibration rig turn out to be smaller than the MPE, meaning that the flowmeter (item) fulfils the specified requirement.

Figure 1: A verification task based on a flowmeter’s Maximum Permissible Error (MPE).
Figure 1: A verification task based on a flowmeter’s Maximum Permissible Error (MPE).

Endress+Hauser’s onsite verification offerings:

Heartbeat verification

• Heartbeat technology verifies the correct function of the measuring device according to the specifications and generates a protocol without process interruption.

• The automatic generated protocol supports the documentation requested by internal and external formalities, laws and standards.

Inline ultrasonic clamp-on verification

• The inline verification is a comparison of the results obtained from the unit under test (UUT) against the inline ultrasonic clamp-on flowmeter.

• A verification certificate which indicates the measured error between both the measurements is generated.

Calibration made easy

According to the VIM, calibration is a procedure to establish a relation between a quantity value given by a UUT and a reference quantity value (ref) obtained by a calibration rig, within its associated measurement uncertainty. The main objective is to check the accuracy of measurements by comparing the device in question with that of a known traceable reference. One fundamental requirement for carrying out a calibration is that the reference system must have traceability to the fundamental units of measurement needed to reproduce the unit flow. Calibration of devices assists:

• The requirements of industry regulators and standards like FDA, IFS, ISO 9000 etc.

• To prevent influence of inaccurate measurements on the quality of the final product.

• To prevent energy or material losses due to improper control.

• To prevent safety issues caused by poor monitoring.

Figure 2: Optimised plant maintenance.
Figure 2: Optimised plant maintenance.

Endress+Hauser’s calibration offerings:

Onsite: portable rigs and buffer solutions

On-site calibration is performed by highly trained engineers. It is convenient and cost effective, and removes the need to send instruments offsite as, keeping downtime to an absolute minimum. It also offers the highest flexibility as calibration can be scheduled according to process demands.

Laboratory

Laboratory calibration services are one-time or repeat contract-based calibrations of customer instrumentation carried out in a facility owned by Endress+Hauser. Calibration services performed in a laboratory have the advantages of the best calibration uncertainty and wide calibration ranges.

Calibration management service

Calibration management service is an optimisation service where Endress+Hauser take day-to-day management responsibility of a customer’s calibration function. Goals of this outsourcing are improving the customer’s plant operations, and securing calibration process compliance to internal and external regulations while reducing its costs.

For more information contact Preston Reddy, Endress+Hauser, +27 11 262 8000, preston.reddy@za.endress.com, www.za.endress.com



Credit(s)



Share this article:
Share via emailShare via LinkedInPrint this page

Further reading:

Acoustic imager detects partial discharge
November 2020, Comtest , Maintenance, Test & Measurement, Calibration
Fluke’s new ii910 Precision Acoustic Imager detects corona, partial discharge and gas, steam and compressed air leaks that can hurt both production uptime and the bottom line.

Read more...
The challenges of online trace measurement
November 2020, Endress+Hauser South Africa , Analytical Instrumentation & Environmental Monitoring
The power industry has extremely high expectations in terms of measuring sensitivity, particularly with regard to the key parameter silica.

Read more...
Safe and efficient water and wastewater treatment
October 2020, Endress+Hauser South Africa , System Integration & Control Systems Design
Depending on the application and the industry, water needs to have different properties to ensure the quality of finished products and the overall safety of production.

Read more...
The Fluke T6 with FieldSense technology
October 2020, Comtest , Maintenance, Test & Measurement, Calibration
FieldSense technology works by providing a capacitive path from a contact point on the back of the tester and the operator’s hand to earth ground.

Read more...
Endress+Hauser strengthens presence in the Middle East
October 2020, Endress+Hauser South Africa , News
Endress+Hauser is strengthening the Group’s presence on the Arabian Peninsula with the establishment of Endress+Hauser Middle East.

Read more...
Preventive maintenance in pulp and paper mills
October 2020, R&C Instrumentation , Maintenance, Test & Measurement, Calibration
The thinking around machine condition monitoring has changed over the years and it is now stated that vibration is the better measurement over temperature, since it gives earlier warning.

Read more...
Leak detection has never been quicker and easier
September 2020, SKF South Africa , Maintenance, Test & Measurement, Calibration
SKF’s TKSU 10 ultrasonic leak detector is able to identify leakages in compressed air and vacuum systems, delivering substantial savings for end-users by reducing inspection time and minimising energy costs.

Read more...
Best practices in measurement
September 2020, Yokogawa South Africa , Maintenance, Test & Measurement, Calibration
Operating envelopes are changing and the impact of these changes on instrumentation must not be underestimated. Instrumentation with a design rating of a particular capacity is heavily stressed when the load is increased.

Read more...
Internal inspection of machinery
September 2020, SKF South Africa , Maintenance, Test & Measurement, Calibration
The TKES 10 Endoscope Series from SKF is designed for fast, easy internal inspection of machinery, delivering significant savings in terms of both time and money to customers and end-users.

Read more...
Parker releases new software tool
September 2020, Parker Hannifin - Sales Company South Africa , Maintenance, Test & Measurement, Calibration
IQANgo enables service technicians or machine owners to connect wirelessly to modules in their machines, allowing them to perform a broad range of actions such as system status checks and troubleshooting errors or alarms.

Read more...